DISINFECTION INTERVALS
WARNING: Do not attempt to open or remove enclosure. There are no user-serviceable internal components.
NOTE: The disinfection process can only be performed by an appropriately trained individual.
PART/ACCESSORY
Outside surface of the conserver and
controls
Oxygen Tubing, nasal cannula
MAINTENANCE
The oxygen conserver does not need maintenance through the expected service life of the product. The device has an
expected service life of 5 years, which is based on 1460 hours/year of use under normal operating and storage conditions.
PREVENTIVE MAINTENANCE
Before attaching the conserver to an oxygen cylinder:
Examine the seal washer and replace if it is damaged or contaminated with dirt, debris, oil or
other contaminants.
Inspect the threads on the cylinder attachment knob, make sure that the knob turns freely through its
full range of adjustment.
To replace the SEAL-100 Seal Washer:
1. Remove the old seal washer by lifting the edge of the brass washer and removing the seal washer from the
oxygen inlet connection, discard old seal washer .
2. Wash and dry hands before handling the new seal washer .
3. Place the new seal washer over the oxygen inlet connection and press it against the body of the devi ce to
seat the seal washer in place.
22
RECOMMENDED DISINFECTION
INTERVAL
Between patients
Do not clean, replace between patients
COMPATIBLE DISINFECTION METHOD
1:5 chlorine bleach (5.25%) and water
solution
N/A