Transcend 365 miniCPAP Auto Quick Guide
Flashing Red X – No Heater Detected
Indicated by a flashing red X. The humidifier will be disabled. Caused by:
This indicates a corruption, most likely caused by moisture in the CPAP. Allowing
the system to dry out may correct the issue. To do this, run the CPAP in Dry Mode.
Note: After drying the system, disconnect and reconnect the power source to ensure the
status is reset.
If assistance is needed to reset an indicator, contact your medical equipment
provider or Somnetics customer service 877-621-9626 or info@somnetics.com
Combination Indicators
The Water and Heater status symbols are not mutually exclusive. It is possible that both
issues may occur and will be shown in combination in the lower right portion of the LCD
screen.
Example:
All manuals and user guides at all-guides.com
Page 14
Flashing red X
Flashing red X