H. Automatic Shutdown
®
ProPex
II automatically shuts down after 5 minutes without use. It is advisable,
however, to manually switch off equipment after measurement by simply pressing
the
key (On/Off).
I.
Maintenance of your ProPex
The device does not contain user serviceable parts. The service and repair
should be provided by factory trained service personnel only.
After each use, all the objects that were in contact with infectious agents should
be cleaned using towels impregnated with a disinfecting and detergent solution
(a bactericidal, fungicidal and aldehyde free solution). Use of chemical agents
may cause damage to the equipment. We recommend to use only a disinfecting
solution which is approved for its efficacy (VAH/DGHM-listing, CE marking, FDA
approval).
Other recommendations:
-
Do not expose ProPex
chemical solutions used for treatment. These chemicals could cause damage,
deform or discolor the device. Be especially careful to avoid spilling formalin
cresol (FC) and sodium hypochlorite as they are quite aggressive. Wipe up
any chemical spills immediately (some chemicals may leave discoloration and
spots even if they are immediately wiped up).
-
Handle the device carefully; do not drop, bump or expose the unit to any kind
of impact or shock. Rough handling could cause significant damage.
-
Do not drop anything on or bang the measuring cable plug after it has been
inserted into the connector.
-
ProPex
®
II must be stored in normal temperature (< 60°C) and humidity
conditions.
Reprocessing procedure:
- The lip clip, the hook and the fork must be sterilized between treatments by
autoclaving at 134°C.
- The measuring cable cannot be autoclaved.
See the section "Disinfection, cleaning and sterilization procedure for the ProPex
accessories" for the detailed procedure.
®