7. Technical data
Classification according to IEC 60601-1
Type of protection against electric shock:
Degree of protection against electric shock:
Degree of protection against harmful
ingress of solids and water:
General specifications
Power supply (voltage):
Frequency:
Maximum current:
Environmental Operating Conditions:
Storage and transport directions:
Manufactured and tested to the following
standards:
Set temperature accuracy:
Dimensions:
Weight:
Guidance and manufacturer's declaration – electromagnetic emissions
The Test Tube Heater is intended for use in the electromagnetic environment specified below. The
customer or the end user of the Test Tube Heater should assure that it is used in such an environment.
Emissions Test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
Complies
IEC 61000-3-3
Class I equipment
Type B
Test Tube Heater: IP64
Power Supply: IP20
100 - 240 VAC
50 - 60 Hz
1.1 A
+15 to +30°C
10 to 90% RH
700 hPa to 1060 hPa
-10 to +50°C
5 to 90% RH
IEC 60601-1: 1988+A1:1991+A2:1995
IEC 60601-1: 2005 + A1: 2012(E)
IEC 60601-1-2: 2014
± 1 °C
Test Tube Heater:
150 x 100 x 28 mm (excluding base/cable)
Power Supply:
121 x 50 x 31 mm (excluding cables)
Test Tube Heater: 0.7 kg
Power Supply: 0.3 kg
Electromagnetic Environment Guidance
The Test Tube Heater uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
The Test Tube Heater is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low voltage
power supply network that supplies buildings used for
domestic purposes.
1-12
English
1