K-FTH-1012 Instructions for Use
1
2
1
3
The temperatures measured should be the set temperature ± 1.0°C. (Standard factory set point is 36.9°C)
6.3 Inspection by an authorised service agent
Inspections at least once a year
For ongoing operational safety of the device, the device must be maintained at regular intervals by an
authorised service agent as per SMM30004. The service agent will verify the performance of the device.
Depending on the duration and frequency of use, the device should be maintained at least once a year.
Failure to adhere to this maintenance schedule will result in the manufacturer refusing liability for the
operational safety of the device.
Authorised service agents
All services such as alterations, repairs, calibrations etc., may only be performed by the manufacturer or
by service agents who are authorised by the manufacturer as per SMM30004.
Liability
The manufacturer is free from all liability for the operational safety of the device if the device has been
wilfully opened and unauthorised persons have performed repairs or alterations on it.
Certification
The device owner should receive a certificate from the service technician for all inspections or repairs.
This certificate states the type and scope of the services rendered, the service date and the name of the
service company. The certificate should be signed.
Technical documentation
If the manufacturer provides you with technical documentation, this does not authorise the user to
perform repairs, adjustments or alterations to the device or accessories.
6.4 Returning the device
If it becomes necessary to return the device, use of the original packaging is required. The
manufacturer does not take responsibility for damage that has occurred during transportation if the
damage was caused by inadequate transport packaging. Please enclose the following information with
the device:
• owner's name
• owner's address
• model type
• Unique Device Identifier of the equipment (printed under barcode on the rear of the device)
• description of the damage or fault
The manufacturer has the right to refuse to carry out repairs if the products it receives are contaminated.
1. Thermometer
2. Modified Test Tube Cap to allow a thermometer to be inserted
3. 6.5 mL of water
1-11
English