1.1 Language
ENGLISH, DEUTSCH, FRANCAIS, ESPANOL
Options
1.2 Application Type (APPL)
STEEL
Options
CONCRETE
1.3 Operation
LEVEL
SPACE
Options
DISTANCE
Operation types
sensor reference point
(flange face)
low calibration point
1.4 Units
Select the units for the Quick Start variables (high and low calibration point, and level, distance,
or space).
mm, cm, m, in, ft
Options
1.5 High Calibration Point (CAL HIGH)
Distance from Sensor Reference to high calibration point: usually process full level (See 1.3
Operation for an illustration).
Range 0.0000 to 100.00 m
Values
1.6 Low Calibration Point (CAL LOW)
Distance from Sensor Reference to low calibration point: usually process empty level (See 1.3
Operation for an illustration)
Range: 0.0000 to 100.00 m
Values
Page EN-12
SITRANS LR460 – QUICK START MANUAL
All manuals and user guides at all-guides.com
Silo construction
Distance to material surface referenced from low calibration point
(process empty level).
Distance to high calibration point (process full level) referenced
from material surface.
Distance to material surface referenced from sensor reference
point.
high
calibration
point
distance
.
space
level
A5E32007360