Télécharger Imprimer la page

Optika Italy IM-3F Manuel D'instructions page 16

Masquer les pouces Voir aussi pour IM-3F:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 68
Centering the ring
Usually this operation is not needed. If
necessary, please proceed with the fol-
lowing steps:
1.
Place a specimen on the stage and focus
it.
2.
Take out the eyepiece from the tube with-
out the dioptric adjustment, and replace it
with the centering telescope (CT). (Fig.40)
3.
Check that the phase ring and the objective
correspond, and that both are steadily set
on a click stop.
4.
Use the CT to focus on the light ring's im-
age ① and the phase contrast ring's image
②. If the light ring's image is not sharp, ad-
just the CT's eyepiece until you can see a
clear image of the light ring.
5.
Adjust the bolts of the two centering holes
in the phase contrast slider using a screw-
driver until the light ring center and the
phase contrast ring center coincide.
6.
The 10x and the 20x phase contrast objec-
tives use the same ring on the phase con-
trast slider. The coincidence of the light ring
center and the phase contrast center must
be verified with both objectives.
(Fig.41; Fig.42)
If the light ring is centered incorrectly,
the contrast will be severely impaired.
The phase ring may need recentering
during and after observation of very
thick specimens.
The phase ring may show an apparent
misalignment if the cover glass is not
flat.
All manuals and user guides at all-guides.com
Page 16
Fig.40
Fig.41
Fig.42

Publicité

loading

Ce manuel est également adapté pour:

Im-3fl4