Moving and storing the scale
– Switch off the scale.
– Unlock the actuating lever of the mea-
suring foot by pressing down the round
knob and folding out the lever.
– Raise the actuating lever to lower the
castor.
– Pull the measuring foot out to the side,
wind up the cable and place the mea-
suring foot in one of the compartments
of the equipment trolley.
– Proceed in exactly the same way to re-
move the remaining three measuring
feet and store them on the trolley.
– Disconnect the mains plug from the
socket and likewise store the mains unit
on the equipment trolley.
– Release the brakes on the equipment
trolley.
You can now move the bed scale to its
next point of use or "park" it somewhere
out of the way until it is required again.
5. Setting damping
Damping is useful for reducing weight determination malfunctions (due to patient move-
ments, for example). The stronger the damping, the more slowly the weight display reacts.
You can set damping to suit requirements.
– Press the FUNC key and keep it de-
pressed for 2 seconds.
FIL and the current setting appear in
the display:
0 = slight damping
fast determination of weight
1 = average damping
normal determination of weight
2 = strong damping
slow determination of weight.
– Press an arrow key if you want to
change the set value.
– Press the FUNC key again to store the
displayed value. The scale is then back
in normal weighing mode.
The settings are retained even once the scale is switched off.
Model 985
GB
33