The fl ow outlet shall not be used for driving any medical device.
Note also that products colour (especially handwheel) might not follow any
gas colour coding.
6. OPERATIONS

6.1. BEFORE USE

6.1.1. VISUAL INSPECTION BEFORE FITTING
EN
Check if there is visible external damage of the fl ow selector (including
product labels and marking). If it shows signs of external damages, re-
move it from service and identify its status.
Visually check if the product is contaminated; if needed, for the fl ow
selector, use the cleaning procedure detailed in this instruction for use.
Ensure the handwheel is set on "0" position (turning anticlockwise).
Ensure that connected hose nipple or humidifi er to the outlet has correct
dimensions.
The fl ow selector is dedicated only for use with the gas specifi ed on its
labelling. Never try to use for another gas.
6.1.2. FITTING TO MEDICAL GAS PIPELINE SYSTEM TERMINAL UNITS
The fl ow selector shall be connected in accordance with suitable type of
terminal unit or with a regulator with an appropriate quick connector, see
appendix nr.2.
6.1.3. FUNCTIONAL TEST
Check that there is a gas fl ow at each setting higher than "0" set by the
handwheel (for instance, by listening of the sound of gas fl ow or check-
ing presence of bubbles in a humidifi er).
Set the handwheel back to the close position – value "0".
If any leakage is detected, use the procedure in chapter 6.3 and return
the product to GCE for service.
6.2. CONNECTION & USE OF OUTLET
6.2.1. LIST OF RECOGNIZED ACCESSORIES
To be connected to the fl ow outlet:
Humidifi er, breathing mask or cannula.
Before connecting any accessory or medical device to the fl ow selector,
always check that it is fully compatible with the product connection fea-
tures & the product performances and there is no negative infl uence to
outlet parameters of the fl ow selector.
6.2.2. FLOW OUTLET CONNECTION
When connecting any accessory to the fl ow outlet make sure that it is not
connected to the patient before operating the product.