Maintenance after each Simulation Session
The following preventive measures are required to ensure longevity of
the SimJunior Simulator.
Power off the SimPad, Link Box and PCs
Charge batteries if necessary
Flush the IV-arm system and drain the IO leg (leave the plug in the
bottom of the IO leg open)
After each session where the IV-arm has been used, flush the IV
arm with 60% Isopropanol alcohol or 70% ethanol
Wipe the skin with a moist cloth
Remove wet clothes or linens
Remove tape residue with a moist cloth or rubbing alcohol
Based on the use of the Simulator, replace modules that are
consumed or damaged
Replace cables or connectors showing visible damage
Servicing should be performed by qualified personnel at regular
intervals
Always perform a service:
If liquids have been spilled in or on the Simulator
After use in dusty environments
Note: For information on and spare parts, see section: Spare
Parts and Accessories.
Before Storage or Shipping
Flush the IV arm with 60% Isopropanol alcohol or 70% ethanol.
Open the drain valve on the bottom of the Simulator's right foot.
Maintenance
12