Note on optional accessory:
-
Temperature probe (K-type) made of V4A tube
application: insertion probe for soft-plastic materials, liquids, gas and air
measuring range: - 196 °C up to 800 °C
dimensions: length = 210 mm, tube length = 120 mm, tube diameter =
3 mm, V4A (P.no. 044121)
Note on replaceable parts:
-
The BENNING MM 6-2 contains fuses for overload protection:
One fuse rated 11 A rapid-acting (1000 V) 30 kA, D = 10 mm, L = 38 mm
(P.no. 10218772).
-
The BENNING MM 6-1/ MM 6-2 is supplied by means of one integrated 9 V
block battery (IEC 6 LR 61).
-
The above mentioned safety cable (tested spare part) are approved in
accordance with CAT III 1000 V/ CAT IV 600 V and for a current up 10 A.
4.

Description of unit

See fig. 1:
The display and operating elements shown in fig. 1 are as follows:
 Digital display, for the measurement reading, bar graph and display for
overrange indication,
 Polarity display,
 Battery display,
 Function key (blue),
 RANGE key, switchover between automatic and manual measuring range,
 Δ/PEAK key, relative value function and peak value storage
 Smart HOLD key,
8 Button (yellow), display illumination,
 Rotating switch, for selecting measuring function.
 Socket (positive
Hz (BENNING MM 6-2)
 COM socket, joint socket for measurement of current, voltage, resistance,
frequency, temperature, capacity, continuity and diode testing,
 Socket (positive), for 10-A range, for currents up to 10 A
 Protective rubber holster
 LED (red) for voltage indicator and continuity test
) The automatic polarity display for DC current and voltage refers to this.
1
5.

General data

5.1 General data on multimeter
5.1.1
The digital display 1 is designed as a 4 digit liquid-crystal indicator with
15 mm digit height and decimal point. The highest value displayed is 6000.
5.1.2
The bar graph display consists of 60 segments.
5.1.3
The polarity indication  functions automatically. Only a polarity con-
trary to the socket definition is indicated, as "-".
5.1.4
The range overload will be displayed with "OL" or "-OL" and sometimes
with an acoustic signal.
Attention: no display or warning by complete overload.
5.1.5
The rotary switch 9 is intended for selecting the measuring function.
5.1.6
The "RANGE" key 5 can be used to change over to the manual meas-
uring ranges and to hide "AUTO" on the display at the same time. By
pressing the key for approx. 2 seconds, the automatic range selection is
activated ("AUTO" on the display).
5.1.7
Press the Δ/PEAK key 6 (relative value function) to store the currently
displayed value and to display the difference (offset) between the stored
measured value and the following measured values. Press the Δ/PEAK
key 6 for 2 seconds to switch over to the "PEAK" function (peak value
storage). The "PEAK" function detects and stores the positive and nega-
tive peak/ crest value (> 1 ms) in the mV/ V AC and mA/ A AC function.
Press the key to call up Pmax, Pmin and the currently measured value
(Pmax, Pmin). Press the key for approx. 2 seconds to switch the device
back to normal operating mode.
5.1.8
"Smart HOLD" - storage of measurement reading: When the "Smart HOLD"
key  is pressed, the measurement reading is stored in the memory.
The symbol "HOLD" appears in the display 1 at the same time. If the
measured value increases by 50 digit above the saved value the change
in measured value is shown by a blinking display and by a signal tone.
(changes of measured values between AC and DC voltage/ current will
not be recognized).
5.1.9
Press the yellow key 8 to activate the display illumination. The device is
switched off automatically after 2 seconds or by pressing the key again.
5.1.10 The function key (blue) 4 serves to select the secondary function of the
rotary switch position.
02/ 2020