Table 9.
Recommended separation distances between portable and mobile RF
communication equipment and the Aerogen Solo nebuliser system that is not life
supporting
This Aerogen Solo nebuliser system is intended for use in an electromagnetic
environment specified in Table 8. The customer or the user of the Aerogen Solo nebuliser
system can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and
the Aerogen Solo nebuliser system as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum
output power of
transmitter W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (w) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
42
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = [1.17] √P
0.12
0.37
1.17
3.70
11.70
Aerogen®
m
80 MHz to 800 MHz
d = [1.17] √P
0.12
0.37
1.17
3.70
11.70
800 MHz to 2.5 GHz
d = [2.33] √P
0.23
0.75
2.33
7.36
23.30