SEFRAM 3516BF Manuel D'utilisation page 69

Pinces wattmé triques 600a et 1000a
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

(11) Flex AC Current (voltage input):
Function
Range(1mV/1A)
ACA
300.0A/3000A
HFR ACA
300.0A/3000A
Peak
420.0A/4200A
INRUSH
300.0A/3000A
Frequency
99.99Hz/999.9Hz
THD
99.9%
Harm H01-H12
99.9%
*The accuracy of sFlex-T is not included.
**ACA <300dgt, add 3 dgt to the accuracy.
- If ACA <30Arms, it will display "rdy" in Harmonic mode.
Trigger level of INRUSH : 1% of current range.
(12) Temperature
Function
Range
-50 °C ~ 399.9 °C
°C
400 °C ~ 1000 °C
-58 °F ~ 751.9 °F
°F
752 °F ~ 1832 °F
- The above specification is assumed at the ambient tempera-
ture stability within ±1°C. In addition, the temperature probe
has to be connected to meter for more than 1 hour in advance.
The meter needs 2 hour for stability for ambient temperature
change more than ± 5°C.
Accuracy*
±(1%+5dgt) (50~500Hz)**
±(1%+5dgt) (50~60Hz)**
±(5%+5dgt) (61~400Hz)**
±(3%+80dgt) (50~500Hz)
±(2%+10dgt) (50/60Hz)
± (0.5%+3dgt) (<500Hz)
±(5%+10dgt)
±(5%+10dgt)
3536BF
Accuracy
± (1% + 3°C)
± (1% + 6°F)
M35X6BF-M-00
34

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

3526bf3536bf

Table des Matières