"HFR"
NOTE :
Peak Hold, Inrush, HZ, Harmonic and HFR mode are only
available in AC mode.
Measuring Active power(W)/Power factor(PF)
1.Single Phase Power Measurement
Step1. Set the rotary switch to the "W" position.
Step2. Connect the Red test lead to the L, and the Black test
lead to the N.
Step3. Press the trigger to open the transformer jaws and clamp
one conductor only, make sure that the jaw is firmly
closed around the conductor.
Step4. Use the MODE button to select the "ACW/DCW/PF"
mode.
600A
+L
600V
-N
M35X6BF-M-00