Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration – electromagnetic immunity
The FT 100 is intended for use in the electromagnetic environment specified below.
The customer or the user of the FT 100 should assure that it is used in such an environment.
Immunity test IEC 60601
test level
3 Vrms
Conducted RF
150 kHz to
IEC 61000-4-6
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
Compli-
Electromagnetic environment – guidance
ance level
Portable and mobile RF communications equipment should be used
no closer to any part of the FT 100, including cables, than the recom-
mended separation distance calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
d = 1.2
3 Vrms
d = 1.2
d = 2.3
where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the
3 V/m
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an elec-
tromagnetic site survey
in each frequency range
Interference may occur in the vicinity of
equipment marked with the following symbol:
229
80 MHz to 800 MHz
800 MHz to 2.5 GHz
a,
should be less than the compliance level
b
.