What to do in the event of measuring errors or operating errors
People suffering from cardiac arrhythmia, vascular constrictions, arteriosclerosis in the extremities
and diabetics should consult their doctor before performing blood pressure self-monitoring because
in such cases the blood pressure levels may be different.
Error
ERR 1: Measuring error
ERR 2: Inflation error
ERR 3: Deflation error
ERR 300
No readings in the display
fields.
Battery symbol
GB-50
Cause
Arm or body movement during
measurement. The cuff is not sitting
properly.
Talking during measurement.
No pressure is being built up in the
system.
Deflation rate too high or too low.
A fault has been detected.
Inflation pressure above 300 mmHg
Battery error
No batteries inserted.
Batteries not inserted properly. Batteries
empty.
Batteries empty.
Remedy
Repeat measurement after 3-5 minutes. Do
not move your arm and do not talk. If the
error occurs again, contact customer service.
Repeat measurement and if the error occurs
again, call the service telephone number.
If the error occurs again, call the service
telephone number.
Repeat measurement after 3-5 minutes. Do
not move your arm and do not talk.
Repeat measurement after 3-5 minutes.
Check batteries
Check batteries and replace if necessary.
Check batteries and replace if necessary.