4.4 Connecting the potential
equalization conductor
To prevent interference with monitoring
devices when used in treatment rooms,
the CPM device must be connected to a
potential equalization system according to
DIN 42801.
5. Setting the treatment values
Warning!
m
Patient hazard –
Before therapy, a test run consisting of
several exercise cycles must be complet-
ed without the patient. Then, repeat the
test run with the patient and check that
the movements do not cause any pain.
Note!
See also 2.2 and 2.3 as well as page VI !
For this purpose, when using devices with
serial numbers > 35,000, connect the equal-
ization conductor to the potential equaliza-
tion terminal (21) of ARTROMOT®-K1 classic
and to the potential equalization connection
of the treatment room.
Observe the requirements of IEC 60601-1 for
potential equalization connections.
5.1 General information on pro-
gramming ARTROMOT®-K1
classic
1. You activate a function by briefly press-
ing a control on the control pendant.
2. You select a treatment value by pressing
the respective control and turning it in
either direction.
You increase a value by turning the
control clockwise towards the + (plus)
symbol and you decrease a value by turn-
ing the control counterclockwise towards
the – (minus) symbol. For the first 5° the
values in the display change in steps of
1°, then the interval changes to 5° so that
you reach the target value faster.
During adjustment of the extension and
flexion angles, the values in the display
change as you turn the control after
pressing it.
When adjusting speed or pause, the in-
formation shown in the display changes
automatically as you press the control.
The selected parameter (speed or pause)
is immediately represented by its symbol
in a large format plus the current value.
(Also refer to 2.2.2).
The current value can be changed by
pressing and turning the control. When
you have set the new value, simply
release the control and the standard
display reappears automatically after
approx. 5 seconds (see also 2.2.1).
3. Subsequently press the START/STOP
button to start therapy.