Single-use component replacement
Recommended personnel: Single-use component
replacement should be performed by a trained
technician.
Equipment and materials required: RENASYS™
inlet O-ring, #1 Pozidriv® screw driver, suction cup
(optional), plastic tweezers, 2 RENASYS odor filter
replacements.
Inlet O-ring (REF 66800603)
Precaution: Do not lubricate O-ring
1.
To ensure the canister seal is secure, the device inlet
port O-ring should be replaced after each patient use.
2.
Clean the device inlet port and carefully slide new
inlet port O-ring onto device inlet port; checking
that any rolling of the O-ring is corrected once it is
located in the groove of the inlet port.
Odor filter (REF 66800061)
1.
To minimize unpleasant odors from RENASYS GO
devices, it is recommended that the device odor filters
be replaced between each patient use.
2.
Turn the device over and lay it on a soft surface to
avoid damage.
3.
Remove Pozidriv® screw holding the rear flap and
open to reveal the odor filter space.
4.
Insert 2 new odor filter elements.
5.
Close the door flap and insert screw, tightening only
until door flap is flush with case.
Precaution: TAKE CARE NOT TO OVERTIGHTEN
SCREW!
Device physical appearance check
Recommended personnel: Device physical
appearance check should be performed by a trained
technician.
1.
Confirm device is unplugged from AC mains power
and placed on a flat surface.
2.
Visually assess device for damage, cracks or
missing pieces.
3.
Confirm buttons and indicator lights below display
screen and display screen itself is undamaged.
4.
Confirm inlet port is undamaged.
5.
Confirm AC mains power inlet is undamaged.
6.
Visually confirm carry strap pins on each side of
device are present and secure.
7.
Visually inspect entire length of power cord for any
fraying, missing insulation, loose or missing plug
blades or earth/ground.
8.
Visually inspect entire length of power supply cord
for any fraying or missing insulation, and plug to
confirm it is not bent, loose or missing.
9.
Rotate device 360° in vertical and horizontal
planes. Confirm device does not have loose
components.
10.
Inspect rear of device and confirm device specification
badge is present and REF and SN numbers are clearly
visible.
11.
Confirm power supply front and rear labels are
clearly visible.
7
Reference
number (REF)
Serial
number (SN)