Error
PROC HEAD OPEN
TURB TOO HIGH
VIAL PRESENT
VIAL CLARITY
8
WATER INGRESS
8
Water drops, puddles or runlets that will not damage the instrument may be in the inner of the
enclosure.
Description
The process head is in the open
position or the process head detector
has a malfunction.
The turbidity reading is higher than
the measurement range of the
instrument (1000 FNU maximum).
There is no vial in the vial
compartment.
The vial or vial compartment is dirty.
There is water in the instrument.
Solution
Turn the process head to the
closed position.
Make sure that the turbidity value
of the sample is within the
measurement range of the
instrument.
Install a vial in the vial
compartment.
Clean or dry the vial and the vial
compartment.
Immediately stop flow to the
instrument. Disconnect the sensor
cable.
The desiccant cartridge can
become hot. Only touch and
remove the desiccant cartridge
when it is at room temperature.