7.

OVERLOAD INDICATOR

DANGER
8.

BASIC FUNCTIONS

8.1.

COMMISSIONING

8.1.1. Activation of batteries

The 3 x 1.5 V "AA" batteries are installed in the factory.
Remove the insulating tab protruding from the battery compartment to enable them.
For future battery changes, refer to Chapter 9.

8.1.2. Starting up the unit

t kN daN lbs ton kg
Display
STOP MAX
of digits

8.1.3. Automatic zero

When the LLX1 is started up, the display will indicate "0" so long as the force measured is less than
10% of the maximum capacity of the unit.
When the load applied to the sensor exceeds the maximum capacity of
the appliance of 15 % (e.g.: a 5 t loaded at 5,75 t) the display unit shows
"HI" indicating an overload condition as shown opposite.
In the event of overload, all stress on the sensor must be completely
relieved and a check made that the appliance returns to zero.
If the appliance shows a stress value, even though tension is not
applied, then it has suffered a permanent distortion. In this case, you
must have the appliance serviced by the manufacturer before continuing
to use it.