Table drifts up.
Table drifts down without weight on table top.
Table has a spongy feeling.
Table drifts down with weight on table top.
Table randomly drifts down with
light weight load applied.
Return To Table of Contents
1-12

Troubleshooting Guide

Problem / Failure
Recommended Action
Make sure that the original table top remains attached
to the base. If so, then replace the gas spring (see
page
1-15).
If the table is difficult to raise and it drops back down
when released, replace the gas spring (see
This is normal as long as it does not drift down.
Adjust the gas spring lock pin for more clearance, so
that there is a little play in the handle.
Replace the gas spring (see
4810-009-001 REV C
page
1-15).
page
1-15).