Preventative maintenance should be performed at a minimum of annually. A preventative maintenance program should
be established for all Stryker Medical equipment. Preventative maintenance may need to be performed more frequently
based on the usage level of the product.
CHECKLIST
_____ All fasteners secure
_____ Top and laminate intact - not damaged
_____ Top edge molding intact - not damaged
_____ Release handle locking and releasing properly
_____ All casters secure
_____ No debris in casters and they roll/swivel freely
_____ Base hood intact - not damaged
CAUTION
Do not apply lubricant of any kind to the gas spring. Lubricant will damage the gas spring.
Unit Serial Number:
Completed by: ____________________________________________ Date: _____________________
A recommended cleaner is a chlorinated bleach solution (5.25% - less than 1 part bleach to 100 parts water). Thoroughly
rinse the solution from the product surface. If other hospital-grade cleaning agents are used, follow the instructions
provided by the cleaning solution's manufacturer. Do not allow cleaning solutions or other fluids to pool on the unit.
Wipe the surface with a soft, dry cloth to remove any moisture.
CAUTION
Failure to properly clean a contaminated surface will increase the risk of exposure to bloodborne pathogens.
Return To Table of Contents
1-10
Preventative Maintenance
Cleaning
4810-009-001 REV C
www.stryker.com