Attachments
Analogue interface
Only use connectors with degree of protection IP 67 (Binder 6-pin,
99-5121-15-06, Heidolph order No. 14-010-006-81) to ensure
adequate protection against splashing water.
Ensure that the connector is properly connected.
Pin assignment analogue interface
PIN 1
+5 V
PIN 2
Start/Stop
PIN 3
0 – 10 V
Direction of
PIN 4
rotation
PIN 5
0 V (GND)
PIN 6
4 – 20 mA
Control voltage
0 – 0.9 V
1 V
10 V
➜
➜
➜
➜
➜
Hei-FLOW Expert/
Control current
Ultimate 01
0 – 4.9 mA
0 rpm
5 mA
5 rpm
20 mA
120 rpm
TTL level 0 V = clockwise rotation, TTL level 5 V = counter-
clockwise rotation.
TTL-pulse 5 V switches the rotation on and off again at the next
pulse. This function is always activated.
For controlling the [direction of rotation] and [rotation start/
stop] functions, a voltage of 5 V is also available at the
interface socket in order to be able to switch with potential-free
contacts.
The analogue interface is detected at a voltage of 0.2 V or a
current of 4 mA;
Hei-FLOW Ultimate: When the interface is detected, the display
[EA]
shows
. As long as the pump drive is controlled via the
analogue interface, the display and all buttons except [Start/
Stop] are inactive.
Hei-FLOW Expert/
Ultimate 06
0 rpm
24 rpm
600 rpm
73