General notes
Transportation
During transport, avoid severe shocks and mechanical stresses that can cause damage
to the device.
Keep the original packaging in a dry and protected place for later use.
Storage
Always store the device in its original packaging. To protect against damage and
unreasonable material aging, store the device in a dry environment that should be as
temperature-stable and dust-free as possible.
Recommended ambient conditions for storage:
•
5 °C – 31 °C up to 80 % rel. humidity
•
32 °C – 40 °C up to 50 % rel. humidity (decreasing linearly)
Acclimatization
After each transport and after storage under critical climatic conditions (e.g. high
temperature difference between inside and outside), allow the device to acclimatize
at room temperature for a minimum of two hours to prevent possible damage from
condensation before putting it into operation at the place of use. If necessary, extend
the acclimatization phase if the temperature differences are very high.
Make all supply connections (power supply, tubing) only after the device has been
acclimatized!
Permissible ambient conditions
The device is designed for indoor use only. Permissible ambient conditions for
operation:
•
5 °C – 31 °C up to 80 % rel. humidity
•
32 °C – 40 °C up to 50 % rel. humidity (decreasing linearly)
•
Maximum height above sea level: 2,000 m
When used in corrosive atmospheres, the service life of the device may be reduced
depending on the concentration, duration and frequency of exposure.
The device IS NOT suitable for outdoor use!
The device IS NOT suitable for use in hazardous areas!
56