Operation
Operation Hei-FLOW Ultimate
Select the pump head
By default, peristaltic pumps Hei-FLOW Ultimate are set to
quick). This setting must be adjusted according to the selected pump head:
➜
Turn off the device.
➜
Press the [Cal] button.
➜
While holding down the [Cal] button, turn the device on again.
–
The display shows the currently set pump head.
➜
Release the [Cal] button and press one of the arrow keys repeatedly until the
desired pump head type is displayed (see table below).
➜
Confirm with [Select].
Displayed value
Pu 0
Pu 1
Pu 2
Pu 3
Pu 4
Pu 8
Pu 9
Pump head
Multi-channel, C 8
Single-channel, SP quick
Single-channel, SP standard, SP vario
not used
not used
Multi-channel, C 12
Multi-channel, C 4
66
PU 1
(pump head type SP