3. Sterilization
For the sterilization method, we recommend the autoclave sterilization method.
Sterilization is required first time you use and after each patient as noted below.
Autoclaving
1 ) Turn off the Power.
2 ) Detach the motor from the Motor Cord. ( Refer to 2-1 Connecting/disconnecting
the Motor and the Motor Cord )
3 ) Clean the surface of the Motor with brush. ( Do not use the metal brush ) , and wipe
it with the cotton moistened with disinfecting alcohol.
4 ) Screw the Motor Cap to the Motor. Put the Autoclave plug to the motor insert.
5 ) Insert into an autoclave pouch. Seal the pouch.
6 ) Autoclavable up to max. 135°C.
ex. ) Autoclave for 20 min. at 121°C, or 15 min. at 132°C.
7 ) Keep the Motor in the autoclave pouch to keep it clean until you use it.
Sterilization at 121°C for more than 15 minutes is recommended by EN13060.
6
Autoclave
Plug
O-Ring
Motor Cap
Fig. 3