beep every second. Change the battery before further use (see Section
9.3 Replacing the Battery).
To switch OFF, press both the ON/OFF button (5) and the MODE/OFF button (6)
simultaneously.
7.2 Monitoring
The Movement Indicator Light (3) flashes with each abdominal movement
during monitoring:-
Go manual-USA_10.14.indd 8
Green
Orange
Orange and
♪
Movement-rate
Warning sounds
(dual-tone beep
every second)
)
Red and full
X
)
audible alarm sounds detected for the selected time period (15, 18 or
Normal abdominal movements.
Very small or weak abdominal movements (eg
very shallow breathing).
Movement rate below 8 movements per minute
If the abdominal movement rate increases to
over 8 abdominal movements per minute, the
Movement Indicator Light will reset to green
(normal) and the warning will stop. To disable
this function, see Section 8.3 Movement-Rate
Warning.
No abdominal movement at all has been
20 seconds). You can select the alarm
activation period (ie the period of no abdominal
movement detection) to be 15, 18 or 20
seconds (see Section 8.2 Alarm Activation
Period). The alarm can be deactivated by briefly
pressing the MODE/OFF button (6) or by
turning Snuza Go! off completely (press ON/
OFF (5) and MODE/OFF (6) buttons simul-
taneously). If the MODE/OFF button is used to
deactivate the alarm, monitoring will continue.
8