Steps for span calibration
1. Press and hold the Tare / Menu-Cal button, and the
Calibration Menu will display.
2. Press Yes to enter the Calibration Menu. To change
the calibration mode, press No until Span Cal (span
calibration) is displayed.
3. Press Yes to begin the span calibration.
4. Calibration masses value will be shown in the screen.
5. After the display shows "Place weights" and "100.0000
g", place weight(s) of 100 g on the pan for calibration.
To change to the calibration point of half full capacity
(e.g. 50 g), press the Function / Mode button.
6. After the screen shows "Place weights" and "50.0000
g", place weight(s) of 50 g on the pan for calibration.
7. Remove the weight(s) from the pan.
PXP Series Balance