Materials
Service life
Mode of operation
Ingress protection
Electromagnetic compatibility - compliance information
This device has been tested and found to comply with IEC 60601-1-2:2014 Ed 4.0 for electromagnetic
compatibility (EMC) as Class B according to CISPR 11:2009. Tests levels are listed in the accompanying tables.
Only use this Philips Avent single/double electric breast pump Premium Plus in combination with the
adapter supplied.
Emissions
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Single electric breast pump
ABS, silicone (motor unit)
Polypropylene, silicone (other parts)
500 hours
Continuous operation
IP22 (motor unit), IP20 (adapter)
Compliance
Group 1
Class B
Complies
Class A Category
Complies
Double electric breast pump
Non-continuous operation
55 min on, 180 min off at 22 °C (72 °F)
(duty cycle 23 %)
20 min on, 90 min off at 40 °C (104 °F)
(duty cycle 18 %)
Electromagnetic environment –
guidance
The device does not use RF energy for its
intended function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
The device is suitable for use in locations in
residential environments and in establishments
directly connected to a low voltage power
supply network which supplies buildings used
for domestic purposes.
English
27