Technical information

Vacuum level
Cycle speed
Motor unit rated input
Mains adapter rated
input
Mains adapter rated
output
Adapter type number
Battery
Power port
Operating conditions
Storage and
transportation conditions
Operating and storage
pressure
Net weight
External dimensions
Device classification
Single electric breast pump
Stimulation: -60 to -200 mbar (-45 to -150 mmHg)
Expression: -60 to -360 mbar (-45 to -270 mmHg)
Adjustable in steps of 20 mbar
Intermittent vacuum
Stimulation: 105-120 cycles/min, depending on vacuum level
Expression: 53-85 cycles/min, depending on vacuum level
5 V d.c. / 1.1 A
External mains adapter 100-240 V a.c.
50/60 Hz
400 mA
5 V d.c. / 1.1 A
S009AHz050yyyy The letters "yyyy" represent the output current from 0100
(1000 mA) to 0180 (1800 mA), increments in steps of 100 mA. The letter "z"
denotes type of plug, it can be: U, V, B, S, C, A, D, K, T, I, E, F. 
Rechargeable battery: 3.6 V 2600 mAh (non-replaceable)
Micro-USB
Temperature: 5 °C to 40 °C (41 °F to 104 °F)
Relative humidity: 15 to 90 % (non-condensing)
Temperature: -25 °C to 70 °C (-13 °F to 158 °F)
Relative humidity: up to 90 % (non-condensing)
For optimal performance of the built-in rechargeable battery, it is
recommended to store the device at room temperature.
700-1060 hPa (<3000 m altitude)
230 g
Motor unit: 145 mm x 95 mm x 45 mm (LxWxH)
Applicable classifications:
Degree of protection against electric shock:
- Class II ME equipment
- Type BF applied part
The breast pump and the adapter are specified as medical electrical system
according to IEC 60601-1.
Double electric breast pump
5 V d.c. / 1.8 A
5 V d.c. / 1.8 A
310 g