Corrective Action - ProMed BDS-700 Instructions D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 38
TROUBLESHOOTING
Error/
Error Display
The display is empty,
even though the Start/
Stop button (A) is
pressed.
The battery
symbol is shown on
the display.
Ekranda arıza kodu
gösteriliyor.
+
Both symbols are
shown on the display.
The display shows
unusual measurement
results.
The device does not
react to any button.
The device does
not react after new
batteries have been
inserted.
If the interference cannot be remedied, please contact the manufacturer before trying
to repair the device on your own.
Cause
The batteries are empty.
The batteries are not inserted
correctly / the polarity (+/-)
is wrong.
The batteries are empty.
You moved yourself or your
arm during measurement.
You talked during
measurement.
The device was moved during
measurement.
The cuff is not attached
correctly.
The rubber tube was not free
or is kinked.
The cuff is not inflated correctly
or air escapes during inflating.
Permanently severe irregular
heartbeat (arrhythmia).
The cuff is not positioned or
closed correctly.
You moved yourself or your
arm during measurement.
You talked during
measurement.
Control panels
Fault due to electromagnetic
fields
33

Corrective action

Pilleri yenileriyle de i tirin.
Insert the batteries correctly.
Replace the batteries by new
ones.
Repeat the measurement and
avoid moving.
Repeat the measurement and
avoid talking.
Repeat the measurement and
avoid moving the device.
Attach the cuff correctly.
Ensure that the rubber tube is
free and not kinked.
Attach the cuff correctly and
check the cuff and the tube for
possible leakages.
Position or close the cuff
correctly.
Repeat the measurement and
avoid moving.
Ölçümü tekrarlayın ve her türlü
hareketi önleyin.
Repeat the measurement and
avoid talking.
Remove the batteries from the
device for at least 5 minutes.
Remove the batteries from the
device for at least 5 minutes.
Ensure that no interfering
sources (mobile phone, etc.)
are in close proximity.
DE
EN
FR
IT
ES
NL
RU
PL
TR
AR

Publicité

Table des Matières
loading

Produits Connexes pour ProMed BDS-700

Table des Matières