OPERATING THE DEVICE

Checking/replacing the battery
To replace/insert the batteries, remove the cover of the battery compartment (H) and
insert 4 batteries 1.5V AA (I). Ensure that the batteries are inserted with their right
polarity (+/-). After inserting, close the battery compartment again.
The Promed BDS-700 checks the charge status of the batteries; the battery status display
(12) shows the low capacity of the battery and the
case, the batteries must be replaced immediately, otherwise the measurement result will
be falsified.
Note:
After a battery change, date and time must be set again.
Caution:
Remove the batteries when the device is not used for a longer time.
Do not use old and new batteries or different battery types at the same time.
Warning: If batteries leak and come in contact with skin or eyes, immediately clean these
DE
areas with plenty of water.
Batteries shall only be used by adults. Keep the batteries out of the reach of children.
It is recommended to use batteries of the same or similar type.
EN
Remove empty batteries from the device.
Dispose of the batteries according to the instructions of the battery manufacturer.
Do not use batteries when they have expired.
FR
Using a power adapter
The device may also be operated using a power adapter. Please ensure that this adapter is
approved for the use with medical products and fulfils the following technical requirements: output
IT
voltage: 6 VDC, output current: 600 mA, DC connector: plus inside, outer diameter 5.5 mm,
inner diameter 2.1mm You can always buy an approved power adapter from Promed.
ES
Note:
When using the power adapter and the batteries at the same time, the batteries will not be
NL
charged. Batteries must always be charged in a separate charger.
RU
Note:
If the power supply is interrupted or the batter status display (12) lights up during measurement,
the measurement must absolutely be repeated.
PL
Connecting the upper arm cuff
To connect the upper arm cuff (G) to the device, insert the tube connector (F) into the tube
connection (E). Please ensure that the connector is inserted into the device without using force.
TR
AR
symbol appears. If this is the
28