14 Instant help
In case of an error in the program process, briefly turn off the balance and disconnect
from power supply. The weighing process must then be restarted from the beginning.
Fault
The displayed weight does
not glow.
The displayed weight is
permanently changing
The weighing result is
obviously incorrect
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
40
Possible cause
• The balance is not switched on.
• The mains supply connection has been interrupted
(mains cable not plugged in/faulty).
• Power supply interrupted.
• Batteries are inserted incorrectly or empty
• No batteries inserted.
• Draught/air movement
• Glass doors not closed
• Table/floor vibrations
• Weighing pan has contact with other objects.
• Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
• The display of the balance is not at zero
• Adjustment is no longer correct.
• The balance is on an uneven surface.
• Great fluctuations in temperature.
• Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
RFE-BA-e-1811