6.0 Description of the LC-Display
Possible display after pressing the start key
If the display is empty, the frequency is out-
side the measurement range.
•
RCD trips correctly, test successful
•
Mains plug incorrect (please reverse) into socket.
•
Voltage off range 190V...255V.
•
Mains / RCD already faulty prior to measurement.
(10.0).
•
RCD has not tripped
•
RCD defect.
•
Range selected too small (wrong range,see pro-
tective switch).
Description of the LC-Display
•
PE is under dangerous high voltage or open
•
Test has been interrupted due to excessive contact
voltage.
•
Potential switched loads falsify measurement re-
sult. In networks with fault grounding, dangerous
contact voltages may occur. In this case, mea-
surement will be interrupted .
•
RCD trippes to fast
•
RCD defect.
•
Incorrectly set test current (excessive range).
•
PE pre-loaded due to running loads.
Other Symbols / Indications:
• Symbol
: Choice of selective RCD.
• Symbol
: Battery strongly discharged.
• For all displays, the contact voltage UL 25V or 50V
is indicated on the display, depending on the pre-
vious selection.
21