SYMPTOM
Lavage Leakage
Cavitron
Insert does not
®
vibrate
8 2 Technical Support and Repairs
For technical support and repair assistance call DENTSPLY Professional CavitronCare
1-800-989-8826 Monday through Friday, 8:00 A.M. to 5:00 P.M. (Eastern Time). For areas outside the U.S., contact your
local DENTSPLY Professional representative.
8
POSSIBLE CAUSE
Punctured Lavage Solution Bag.
Porous Lavage Tubing Assembly.
Cracked Handpiece Adapter
Cracked Steri-Mate Handpiece.
Incorrect tubing size.
O-ring worn or missing on Cavitron
Insert.
O-ring worn or missing on Cavitron
Handpiece Cable.
O-ring worn or missing on handpiece
adapter.
Insert defective.
Steri-Mate Handpiece defective.
Steri-Mate Handpiece Adapter
defective.
CORRECTIVE ACTION
Replace lavage solution bag.
Replace tubing assembly.
Replace the handpiece adapter.
Replace the Steri-Mate® Handpiece.
Check the tubing for snug fit on the adapter
barb. If tubing is loose, replace it with the
correct tubing (.110 inch / 2.794 mm I.D.).
Replace the O-ring. Green (12-pack)
®
Item #62605, Black (12-pack) Item #62351
Replace the O-ring. Item #79357.
®
Replace the O-ring. Item #79357.
Replace insert with a new Cavitron
Replace the Steri-Mate Handpiece with a
known working handpiece.
Remove the Adapter and plug the Handpiece
directly into the handpiece cable. If the
Cavitron
Insert scales properly, replace the
®
Adapter.
Factory Certified Service at
SM
Insert.
®