Transport
The gas detector contains a lithium-ion battery. During
transport, in particular air transport, of the gas detector,
adhere to the relevant safety regulations and markings
for lithium batteries.
8
Storage
Dräger recommends storing the gas detector in the
charging cradle.
The sensors are supplied with power even when the
gas detector is switched off to ensure a faster warm-up
time when switching on.
When storing the gas detector outside of the charging
cradle, power supply to the sensors is cut off
automatically after 21 days. When the gas detector is
switched on again, there is an increased warm-up time.
9
Disposal
This product must not be disposed of as house-
hold waste. This is indicated by the adjacent sym-
bol.
You can return this product to Dräger free of
charge. For information please contact the
national marketing organizations or Dräger.
Batteries must not be disposed of as household
waste. This is indicated by the adjacent symbol.
Dispose of batteries at battery collection centers
as specified by the applicable regulations.
10 Technical data
10.1 Gas detection instrument
Ambient conditions:
during operation and stor-
age
Degree of protection
X-am
-20 to +50 °C
700 to 1300 hPa (measur-
ing function)
800 to 1100 hPa (use in
explosion-hazard areas)
10 to 90 % (briefly up to
95 %) rel. hum.
1)
IP 67
®
3500
Alarm volume
Orientation
Gas detector storage time 1 year
Sensor storage time
Battery
Dimensions
Weight (with pump)
Update interval for the dis-
play and signals
1) Tested without the pump or calibration adapter.
Operating time in normal conditions (diffusion
1)
mode)
:
With CatEx and 3 EC sen-
sors
With 3 EC sensors
1) The nominal operating time of the gas detector at an ambient
temperature of 20 to 25 °C, 1013 mbar, alarms less than 1 %
of the time, power-save mode display activated. The actual
operating time varies depending on the ambient temperature
and pressure and battery and alarm conditions.
In pump mode: If the gas detector is used
permanently in pump mode, the operating time is
reduced by approx. 2 h.
Transport
Typically 100 dB (A) at a
distance of 30 cm
Any
The ambient conditions
and the storage time for
sensors in their original
packaging correspond to
those of the gas detector
Lithium-ion battery,
rechargeable, 4.8 V,
6.0 Ah, 28.8 Wh, 250 g
approx. 179 x 77 x 42 mm
(H x W x D)
Typically 550 g, depend-
ing on which sensors are
equipped, without trans-
port harness
1 s
typically 24 h
typically 120 h