1. Connect the hose (3 mm internal diameter) with the
dust and water filter to the inlet spout (see figure M
on page 3) of the pump adapter.
2. Mount the pump adapter to the gas detector. Make
sure that both guide pins are in the correct grooves.
Check that the pump adapter is mounted
correctly. If the pump adapter is mounted correctly,
the leak test starts automatically. If the leak test
does not start, the gas detector is not ready for use.
Avoid bending the pump adapter.
The gas detector automatically switches to pump
mode as soon as the pump adapter is mounted.
The leak test starts automatically.
Dräger recommends performing the leak test
directly before use with the connected probe (hose
probe, rod probe) so that leaks can be detected in
the entire intake system.
3. When the leak test is displayed, the suction inlet on
the probe or hose closes within 60 s and remains
closed until the leak test is complete.
4. Open the suction inlet.
– Leak test successful: Monitoring starts. Observe
the purging times!
– Leak test failed: Inspect the probe, hose and
adapter and repeat the leak test.
5. Place the probe or the end of the hose on the
sampling location.
The temperature at the measurement location may
deviate from the temperature in the gas detector, which
may influence the measured value display. The correct
functioning of the temperature correction can only be
guaranteed on the gas detector.
!
To stop measuring with the pump:
1. Loosen the screw on the pump adapter.
2. Remove the pump adapter.
The pump is flushed and the gas detector
automatically switches to diffusion mode.
4.11 Configuring the device settings
Additional settings can be changed using the Dräger
CC-Vision PC software.
To open the device settings:
1. Select
in measuring mode and confirm the
dialog.
2. Sign in with the required user level, if necessary.
3. Select and confirm Settings.
4.11.1 Activating day or night mode
1. Open the device settings.
2. Select and confirm Night mode / Day mode.
4.11.2 Changing the device language
1. Open the device settings.
2. Select Language.
3. Select and confirm the desired language.
4.11.3 Set date and time
1. Open the device settings.
2. Select Date & Time.
3. Select Set date format and then select and confirm
the date format.
4. Select Set date and then set and confirm the date.
5. Select Set time and then set and confirm the time.
The user must manually switch between summer
and winter time.
When using the X-dock maintenance station,
automatic time synchronisation is possible.
®
X-am