Télécharger Imprimer la page

Renfert Twister Mode D'emploi page 32

Masquer les pouces Voir aussi pour Twister:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 37
Error code
Err. 3
Err. 4
Err. 10
For all other error codes:
Note error code
Have the device repaired.
Quote the error code to the repair service
The following information is intended to
help purchasers ensure safe operation of
this appliance in their laboratory.
Instruct operators on the appli-
cation, possible hazards during
use and operation of this appli-
ance based on the user informa-
tion.
Ensure that this user information is avail-
able to operators.
EN
Cause
The vacuum is built-up
too quickly due to the
fact that the suction
system is blocked.
Motor fault.
Too much material
Mixing time too long,
the material has already
begun to set.
Information for the purchaser
Have filter cleaned or changed
(Sec. 3.3).
Ventilate bowl and remove (see
Sec. 2.6.1, points no. 2 and 3)
Have the device repaired.
Ventilate bowl and remove (see
section 2.6.1, points no. 2 and 3)
Only fill the bowl to the maximum
fill level.
This limit is valid for powders and
liquids measured before mixing.
Ventilate bowl and remove (see
section 2.6.1, points no. 2 and 3)
Select shorter mixing times.
A. Application
A.1 Correct use of the appliance
The Twister / Twister venturi vacuum
mixer is designed solely for homogene-
ous, bubble-free mixing of dental casting
and modelling materials such as plasters,
investments, and silicones.
- 10 -
Solution

Publicité

loading

Ce manuel est également adapté pour:

Twister venturi18261827