GB
Maintenance and cleaning
Daily maintenance:
The temperature in the vaccine
compartment must always be monitored
on the thermometer every day. Please note
that the vaccine in the appliance must not
freeze.
Weekly maintenance:
During normal use, water can accumulate
at the bottom of the appliance; remove with
a cloth or through the drainage hole.
Wipe off water droplets on the inside wall at
the same time.
Check if lid gasket is sealing tight to the top
frame when the lid is closed. A tight seal-
ing lid reduces accumulation of water and
formation of rime significantly.
Monthly maintenance:
Clean the grille on the right side of the re-
frigerator once each month. Clean vaccine
compartment with lukewarm water and mild
detergent Clean outside of refrigerator with
lukewarm water and mild detergent
Yearly maintenance:
Electric connections and components are
to be checked and cleaned once a year or
more if necessary.
Cleaning
Disconnect the power supply before
cleaning. The best way to clean the
appliance is by using luke warm water with
a small amount of unscented detergent.
Never use cleaning agents that scour. Use
a soft cloth. Rinse with clean water and dry
thoroughly. It is important to prevent water
from running into the control panel.
The gasket around the lid must be cleaned
regularly to prevent discolouration and
prolong service life. Use clean water. After
cleaning the sealing strip, check that it
continues to provide a tight seal.
If the appliance is not being used for any
period of time, switch off the appliance,
disconnect the power supply, empty the
appliance, clean the inside, and leave the
lid open to allow air circulation and prevent
bad odor.