6.3 Mounting the pipette clamp
6.4 Checking the leak tightness
1. Insert the pipette.
2. Fill the pipette with water.
3. Hold the pipette vertically.
4. Observe the pipette outlet for approx. 30 seconds.
Do not touch the pipette. Do not press the control buttons.
No water may be allowed to escape.
5. If water escapes, disassemble and carefully reassemble the pipette clamp.
6.5 Replacing
the rechargeable battery
Proceed as follows:
1. Slide the battery compartment lid open.
2. Remove the battery.
3. Insert a new battery.
1. P ush the seal with the groove up into
the filter adapter.
2. Push the wide opening of the mem-
brane filter into the narrow opening
of the pipette adapter.
3. G uide the aspirating cone over the
pipette adapter and turn it until it
engages.
ErgoOne
®
FAST · Manual · 27