SITRANS LVS100/200 is normally mounted into the vessel top for full detection, or
through the tank wall at the detection level, for full, demand, or empty detection.
*
max. deviation
from vertical
max. length L
α
4 m
45º
1.2 m
>45º
0.6 m
7ML19985FT63
SITRANS LVS100/200 – INSTRUCTION MANUAL
Cable
extension
*
version
L
α
Rigid extension with
sliding sleeve
Protection in case of
high material loading.
Use angle mounting for flowing
material only.
If angle mounting is required
with high material loading,
customer-supplied protection
from falling material must be in
place.
Position tines vertically
to avoid product build-
up: use a 50 mm open-
end wrench to turn the
process connection
until the tine
orientation marking
faces up or down.
Cable gland faces
downward to avoid
water penetration.
Tine orientation
marking facing
sideways.
Page EN-9