Name
Rhythm leads settings
Print settings
Export settings
ECG display format
settings
ECG display color
settings
Filter settings
1) Single rhythm lead: one of the twelve leads is selected as rhythm lead. Single lead
rhythm is 1R lead in the format of 3 × 4 + 1R, 6 × 2 + 1R and 1x1.
2) Three rhythm leads: three of the twelve leads are selected as rhythm leads. The
three rhythm leads are the 3R leads in the format of 3 × 4 + 3R.
1) Including horizontal printing report format and vertical printing report format. The
horizontal printing format includes horizontal 12x1, horizontal 6x2, horizontal 6x2 +
1R, horizontal 3x4 + 1R, horizontal 3x4 + 3R, and horizontal rhythm lead; the vertical
printing format includes vertical 12x1, vertical 6x2, vertical 6x2 + 1R, vertical 3x4 + 1R,
vertical 3x4 + 3R, and vertical rhythm lead.
2) Print report with ECG grid: enable this item, and the report will be printed with ECG
grids.
3) Print report with color grid: enable this item, and the report printed will be colorful
4) Grid is represented by dots: enable this item and the report will be the grid in the
form of dots.
5) Enable preview function before print: enable this item to display preview interface
when printing report manually.
6) Auto-half gain in case of high QRS: if this option is selected, the high gain waveform
of chest lead in the printing report will be halved automatically.
7) Auto print: enable this item to stop ECG data acquisition and auto print report.
8) Thick grid / thin grid: set the color of reporting ECG grid line, which can be modi-
fied manually.
9) Report title: set the title of the report.
10) Enable this item, there will be one more page in the report filled with detailed infor-
mation of average template.
1) Export format: PDF, JPEG, BMP, FDA-XML(optional), EM-XML, SCP (optional), BKG
(optional), DICOM.
2) Export report display format: horizontal format includes horizontal 12x1, horizon-
tal 6x2, horizontal 6x2 + 1R, horizontal 3x4 + 1R, horizontal 3x4 + 3R, and horizontal
rhythm lead; vertical format includes vertical 12x1, vertical 6x2, vertical 6x2 + 1R, verti-
cal 3x4 + 1R, vertical 3x4 + 3R, and vertical rhythm lead.
1) Display format: there are six display formats in 12 lead mode: 12x1, 6x2, 6x2 + 1R, 3x4
+ 1R, 3x4 + 3R, 1x1.
2) Display all paper speed: if this option is checked, all paper speeds will be dis-
played, and all paper speeds can be switched in the main interface; if this option is not
checked, only common paper speeds will be displayed, and only common paper speed
can be switched in the main interface.
3) Paper speed: set the default paper speed of the ECG waveform in the main inter-
face. All paper speeds include 5 mm / s, 6.25 mm / s, 10 mm / s, 12.5 mm / s, 25 mm / s
and 50 mm / S; common paper speeds include 12.5 mm / s, 25 mm / s and 50 mm / s.
4) Display all gains: enable to display all gains, display all gains, and all gains can be
switched in the main interface ; disable to display all gains, display only common gains,
and only common gains can be switched in the main interface.
5) Gain: set the default gain of the ECG waveform of the main interface. All the gains
include 1.25 mm / mV, 2.5 mm / mV, 5 mm / mV, 10 / 5 mm / mV, 10 / 5 mm / mV, 20 /
10 mm / mV, 20 mm / mV, 20 mm / mV, 40 mm / mV and AGC (automatic gain). Com-
mon gains include 5 mm / mV, 10 mm / mV and 20 mm / mV.
The background and waveform of the main interface support multiple color selection.
1) AC Filter: provide 50Hz, 60Hz and off 3 options. The purpose of AC filter is to elim-
inate the interference of AC power supply. The AC filter value should be set according
to the natural frequency of the national AC power supply.
2) Baseline filter: 7 options are provided: 0.05Hz, 0.16 Hz, 0.25 Hz, 0.32 Hz, 0.5 Hz, 0.67
Hz and off. Baseline filter makes the baseline of ECG waveform regress rapidly, and the
larger the filter value, the faster the baseline regression.
3) EMG filter: it provides 7 options including 20Hz, 25Hz, 30Hz, 35Hz, 40Hz, 45Hz and
off. The smaller the filter value is, the stronger the ability of eliminating EMG interfer-
ence is.
4) Low pass filter: there are four options: 75Hz, 100Hz, 150Hz and off.
www.spengler.fr