6. Care, Maintenance, Transport, Conversion
The ARTROMOT®‑S3 is suitable for re‑use
if the following points are observed.
6.1 Care/re-use
Warning!
m
Shock hazard –
Unplug the device from the power line
before cleaning.
Shock hazard, equipment damage –
Liquids must not enter the device or the
programming unit.
• The ARTROMOT®‑S3 must be disinfected
by wiping down with a disinfectant. Thus,
it complies with the special hygiene stan‑
dards for medical technical equipment.
• The enclosure and removable armrest
assemblies can be cleaned with com-
monly used disinfectants and mild
household detergents.
• Only use a damp cloth to wipe the CPM
device down.
Warning!
m
Patient hazard –
patient contamination
− Before using the device on another
patient, be sure to clean and disinfect
it according to the instructions given
here.
Note!
The manufacturer recommends using
only special disinfectants approved for
medical devices and with the characte‑
ristics specified below under "Caution".
Disinfection with the following disin‑
fectants and in compliance with the
instructions for use has been evaluated
and approved by the manufacturer in
the risk assessment:
– DESCOSEPT AF lemon
(Part no.: 00‑311L‑xxx)
BAuA Reg.No.: N‑55153,
Manufacturer: Dr. Schumacher GmbH
Contact time: 2 minutes minimum
Caution!
m
Equipment damage –
− The plastic material used is not
resistant to mineral acids, formic acid,
phenols, cresols, oxidants and strong
organic or inorganic acids with a pH
value below 4.
– Use only clear disinfectants to prevent
discoloration of the device.
– Do not expose the CPM device to
strong ultraviolet radiation (sunlight)
and fire.
6.2 Maintenance
(fuse replacement)
Check before each use
Visually inspect the device for signs of
mechanical damage before each use.
If you detect damage or malfunctions that
may impair the safety of the patient or
of the operator, have the device repaired
before using it.