12. IEC 60601‑1‑2 and IEC 60601‑1‑11
The ARTROMOT®-S3 device is subject to
particular precautions regarding electro‑
magnetic compatibility (EMC). The device
must be installed and put into service
strictly in compliance with the EMC
directives set forth in the accompanying
documents.
Portable and mobile RF communication
systems may affect the ARTROMOT®-S3.
The ARTROMOT®-S3 should not be
used adjacent to or stacked with other
equipment. If adjacent or stacked use is
necessary, ARTROMOT®-S3 should be
observed to verify normal operation in the
configuration in which it will be used.
If you detect damage or malfunctions that
may impair the safety of the patient or
of the operator, have the device repaired
before using it.
WARNING!
m
Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
12.1 Electromagnetic emissions
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
ARTROMOT®-S3 is intended for use in the electromagnetic environment specified below.
It is the responsibility of the customer or user to ensure that the ARTROMOT®-S3 unit is
used in such an environment.
Emissions test
RF emissions to CISPR 11
RF emissions to CISPR 11
Harmonic emissions to
IEC 61000‑3‑2
Voltage fluctuations/flicker
emissions to IEC 61 000‑3‑3
If it is necessary to replace assemblies or
cables, only the manufacturer's original
parts must be used to ensure continued
compliance with EMC requirements after
repair. This requirement applies to the
power supply unit, cables and cable lengths,
drive unit consisting of the motor and the
control system, the programming unit incl.
the coiled cable and the connector.
Regarding the electromagnetic compatibi‑
lity, there is no maintenance work required
during the expected service life of the
product.
The product designation ARTROMOT®-S3
used in the text below includes all product
variants.
Compliance
Electromagnetic environment –
guidance
ARTROMOT®-S3 uses RF energy
Group 1
only for its internal function.
Therefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.
ARTROMOT®-S3 is suitable for use
Class B
in all establishments, including do‑
complies with class
mestic and those directly connected
A requirements
to the public low‑voltage power
supply network that supplies buil‑
complies
dings used for domestic purposes.
248