Télécharger Imprimer la page

ARTROMOT S3 Mode D'emploi page 254

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 84
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
ARTROMOT®-S3 is intended for use in the electromagnetic environment specified below. It
is the responsibility of the customer or user to ensure that the ARTROMOT®-S3 unit is used
in such an environment.
Immunity test
IEC 60601
test level
Home healthcare
Conducted RF to
6 V
IEC 61000‑4‑6
0,15 MHz-80 MHz
80% AM at 1 kHz
Radiated RF to
10 V/m
IEC 61000‑4‑3
80 MHz‑2,7 GHz
80% AM at 1 kHz
Proximity fields
27 V/m
from RF wireless
380-390 MHz
communications
PM 18 Hz
equipment to
28 V/m
IEC 61000‑4‑3
430-470 MHz
FM +/- 5 kHz
1 kHz sine
9 V/m
704-787 MHz
PM 217 Hz
28 V/m
800-960 MHz
PM 18 Hz
28 V/m
1,7-1,99 GHz
PM 217 Hz
28 V/m
2,4 GHz-2,57 GHz
at PM 217 Hz
9 V/m
5,1 GHz-5,8 GHz
at PM 217 Hz
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the ARTROMOT®-S3 exceeds the applicable
ARTROMOT®-S3 unit should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or
relocating the ARTROMOT®-S3 unit.
Compliance
Electromagnetic
level
environment – guidance
Home healthcare
6 V
0,15 MHz-80 MHz
80% AM at 1 kHz
10 V/m
80 MHz‑2,7 GHz
80% AM at 1 kHz
27 V/m
380-390 MHz
PM 18 Hz
28 V/m
430-470 MHz
FM +/- 5 kHz
1 kHz sine
9 V/m
704-787 MHz
PM 217 Hz
28 V/m
800-960 MHz
PM 18 Hz
28 V/m
1,7-1,99 GHz
PM 217 Hz
28 V/m
2,4 GHz-2,57 GHz
at PM 217 Hz
9 V/m
5,1 GHz-5,8 GHz
at PM 217 Hz
250