ANI Monitor V1 - Monitorage continu de l'analgésie patient
REUTILISER UN CAPTEUR DEJA UTILISE SUR UN AUTRE PATIENT PEUT
ENGENDRER UN RISQUE DE CONTAMINATION CROISEE.
SI LE PATIENT DEVELOPPE UNE ERUPTION CUTANEE OU D'AUTRES
SYMPTOMES INHABITUELS, RETIREZ LES ELECTRODES. IL EST IMPORTANT
D'APPORTER UN SOIN PARTICULIER AUX PATIENTS SOUFFRANT DE
PROBLEMES DERMATOLOGIQUES.
NE PLACEZ JAMAIS LES ELECTRODES SUR DES BLESSURES.
RISQUE D'ELECTROCUTION :
NE PAS DEBRANCHER LE CORDON D'ALIMENTATION AVEC DES MAINS
-
MOUILLEES OU SIMPLEMENT HUMIDES
-
NE
PAS
ENLEVER
FONCTIONNEMENT OU SA MISE SOUS TENSION
-
LE FABRICANT DE CET APPAREIL A VERIFIE QUE LE COURANT DE
FUITE DE TERRE ET LE COURANT DE SECURITE PATIENT ETAIENT
INFERIEURS AUX LIMITES SPECIFIEES PAR LES NORMES DE SECURITE
APPLICABLES. LA VERIFICATION DE CES COURANTS EST UNE MESURE
DE
SECURITE
REGULIEREMENT ET SURTOUT APRES UNE INTERVENTION TECHNIQUE
D'UN SERVICE BIOMEDICAL.
-
APRES TOUTE PROJECTION DE SANG OU DE LIQUIDE, LE COURANT DE
FUITE DOIT ETRE RETESTE AVANT TOUTE REUTILISATION DU
MATERIEL
PRENDRE LES MESURES DE PRECAUTION COURANTES POUR EVITER TOUT
CONTACT AVEC LE SANG OU D'AUTRES MATIERES A RISQUE DE
BIOCONTAMINATION. LE MATERIEL CONTAMINE DOIT ETRE TRAITE EN
MD/PRD/IN16.ANIV1 V.15 – 29 SEP 2021
LE
CAPOT
DU
QUE
L'ETABLISSEMENT
MONITEUR
PENDANT
DOIT
ENTREPRENDRE
SON
9