7.5
Wipe disinfection for electrical devices without sterilization
Phase
Step
I
Wipe disinfection
RT:
Room temperature
Phase I
►
Remove any visible residues with a disposable disinfectant wipe.
►
Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
►
Observe the specified application time (1 min minimum).
7.6
Inspection, maintenance and checks
►
Inspect the product after each cleaning and disinfecting cycle to be
sure it is clean, functional and undamaged.
►
Set aside the product if it is damaged.
8.
Maintenance
The display "Battery replacement recommended" on the Acculan 4 charger
GA320 (see TA014534/TA014535) indicates when the battery should be
replaced due to reduced capacity.
No other maintenance measures must be performed on the battery.
For corresponding services, contact your national B. Braun/Aesculap
agency, see Technical Service.
9.
Troubleshooting
►
Replace product: have defective products repaired by Aesculap Techni-
cal Service, see Technical Service.
Issue
Battery not working
Battery quickly empty
D
t
Conc.
[°C/°F]
[min]
[%]
≥1
RT
-
Detection
No signal tone when inserting the
battery into the machine
Symbol "Battery replacement recom-
mended" illuminates at the end of
the charging process on the charger
GA320
Water quality
Chemistry
-
Meliseptol Wipes sensitive
Cause
Battery defective
Battery cells at end of life
Remedy
Replace product.
Have battery cells replaced by
Aesculap Technical Service.