Inspect the new product after removing its transport packaging and
prior to first use to ensure it is in good working order; charge with char-
ger GA320 and verify functionality.
To prevent damage caused by improper setup or operation, and to not
compromise the manufacturer warranty and liability:
– Use the product according to these instructions for use.
– Follow the safety and maintenance instructions.
– Only combine Aesculap products with each other.
– Follow the application advisories acc. to standard, see Extracts from
relevant standards.
Ensure that the product and its accessories are operated and used only
by persons with the requisite training, knowledge, or experience.
Keep the instructions for use accessible for the user.
Always adhere to applicable standards.
Do not use the product if it is damaged or defective.
Set aside the product if it is damaged.
Store product at room temperature (Maximum recommended duration
of storage is 365 days; keep out of direct sunlight).
Keep product dry.
Observe the instructions for use for the corresponding Acculan
machines.
4.
Product description
4.1
Scope of Supply
Art. no.
Designation
GA346
Acculan 4 Li-Ion battery short
TA014537
instructions for use Acculan 4 Li-Ion battery short
4.2
Components required to operate the device
Acculan 4 charger GA320
Acculan 4 or Acculan 3Ti machine:
– GA340
– GA341
– GA344
– GT666
– GA670
– GA671
– GT668
– GA675
4.3
Operating principle
The Acculan 4 Li-Ion battery supplies the Acculan machines with short
handle with voltage. The battery is charged in the Acculan 4 charger
GA320. The charging process of a completely discharged battery can take
up to 120 min.
The speed and rotational direction are set depending on the trigger setting
of the machines:
for saws and Dermatome: speed clockwise
for drills: speed clockwise, counter-clockwise, oscillation and pilgrim
step
5.
Preparation
Prior to use, inspect the product and its accessories for any visible dam-
age.
Use the products and their accessories only if they are in perfect tech-
nical condition.
6.
Working with the device
6.1
System set-up
See instructions for use for the respective Acculan machine and the
Acculan 4 charger GA320.
6.2
Functional test
See the instructions for use for the respective Acculan machine.
6.3
Safe operation
Risk of injury and material damage due to inappro-
priate use of batteries!
WARNING
Charge battery only in the Acculan 4 charger
GA320.
Insert battery correctly (Acculan machine and
charger).
Risk of injury and material damage due to damaged
batteries!
Check batteries for any damage before charging.
WARNING
Do not use or recharge damaged batteries.
For proper loading and removal of the battery from the Acculan
machine, see the instructions for use for the respective Acculan
machine.