Smooth Key; Set-Up Function (Instrument Confi Guration) - Chauvin Arnoux C.A 6549 Notice De Fonctionnement

Masquer les pouces Voir aussi pour C.A 6549:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 1
4.4.

/ SMOOTH KEY

The SMOOTH secondary function activates / deactivates an insulation measurement digital fi lter. It
affects only the display (which is smoothed), not the measurements.
This function is useful if the insulation values displayed are very unstable.
The fi lter is calculated as follows:
RSMOOTH = RSMOOTH + (R – RSMOOTH) / N
Since N is set to 20, the time constant of this fi lter is approximately 20 seconds.
4.5. SET-UP FUNCTION (INSTRUMENT CONFIGURATION)
This function, located on the rotary switch, can be used to change the confi guration of the instrument
by accessing directly the parameters to be modifi ed.
Turning the rotary switch to SET-UP gives you access to the menu of all modifi able parameters. Select
the parameter to be modifi ed and its value using the
4.5.1. SET-UP MENU
SET-UP
Instr.Nr. 700016
SW Version 1.8
Display Contrast
Alarm Settings
Adjust Voltage 1
Adjust Voltage 2
100 V
Adjust Voltage 3
250 V
Timed Run (h:m)
Sample Time (m:s)
DAR (s/s)
30/60
SET-UP
Instr.Nr. 700016
SW Version 1.8
Clear Memory
V Disturbance / V Output
Buzzer
Power Down
BaudRate
9600 / RS 232
Units
Europe
Date (d.m.y)
27.04.2009
Time (h:m)
SET-UP
Instr.Nr. 700016
80
PI (m/m)
Set Step Function 1
50 V
Set Step Function 2
Set Step Function 3
Temperature Unit
0:10
Defaut Probe Temperature
Rc Reference Temperature
0:10
ΔT for R/2
3%
On
On
10:21
85
,
,
and
keys.
Instr.Nr. 700016
SW Version 1.8
Calculate ΔT from Memory
1.0/10
Maximum Output Voltage
Set Defaut Parameter
Clear Memory
V Disturbance / V Output
Celsius
23°C
Buzzer
Power Down
40°C
10°C
BaudRate
SET-UP
SW Version 1.8
5100V
3%
On
On
9600 / RS 232

Publicité

Table des Matières
loading

Table des Matières