Chapter 4 - Technical Specification
4.3.1 Guidance and manufacturer declaration - electromagnetic
emissions
TABLE 10: List of electromagnetic emission compliance
IEC60601
Emissions test
compliance
test level
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A
IEC61000-3-2
Voltage fluctuations/
Pst
flicker emissions
Dmax
IEC61000-3-3
4.3.2 Guidance and manufacturer declaration – electromagnetic
immunity
TABLE 11: List of electromagnetic immunity compliance
IEC60601 compliance test
Immunity test
level
Electrostatic
±6 kV contact
discharge
±8 kV air
IEC61000-4-2
Electrical fast
±2 kV for power supply lines
transient/burst
±1 kV for input / output lines
IEC61000-4-4
Surge
±1 kV line to line
IEC61000-4-5
±2 kV line to earth
Page 56 of 65
English
Electromagnetic environment-guidance
The P200T uses RF energy only for its internal function. Therefore,
its RF emissions are very low and are
interference in nearby electronic equipment.
The P200T is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Electromagnetic environment-guidance
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
likely to cause any
NOT
Part Number: G102748/9GWE
Copyright 2017, Optos plc. All rights reserved.